Remote control interfaces and protocols
R&S
®
NRP Series
49Getting Started 1419.0170.02 ─ 16
The default port for socket communication is port 5025.
Example:
A power sensor has the IP address 10.111.11.20; the valid resource string using
VXI-11 protocol is:
TCPIP::10.111.11.20::INSTR
The DNS host name is nrp18sn-100001; the valid resource string is:
TCPIP::nrp18sn-100001::hislip0 (HiSLIP)
TCPIP::nrp18sn-100001::inst0 (VXI-11)
A raw socket connection can be established using:
TCPIP::10.111.11.20::5025::SOCKET
TCPIP::nrp18sn-100001::5025::SOCKET
Ethernet interface