EasyManua.ls Logo

R&S ZC Series - Basic Operation; Required Equipment

R&S ZC Series
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Basic operation
R&S
®
ZCxxx
26Getting Started 1177.5156.02 ─ 05
6 Basic operation
This chapter describes how to configure the standard setup with an R&S ZNA
network analyzer and two frequency converters R&S ZC330, for 2-port transmis-
sion measurements.
Measurements using other converters of the R&S ZCxxx family are performed in
an analogous way.
6.1 Required equipment
In general, the standard setup with n frequency converters requires an R&S ZNA
with at least n test ports, equipped with the following hardware options:
Direct IF access (R&S ZNA-B26)
Converter LO output (R&S ZNA-B8)
The standard RF cabling and the required accessories are described in Chap-
ter 5, "RF connections", on page 20.
To calibrate the measurement setup, the following equipment is required:
For power calibration and leveling, a power meter that covers the measured
frequency range:
VDI Erickson PM5 or PM4 (with suitable taper) for waveguide ports
Alternatively, R&S NRP75TWG or R&S NRP110TWG for waveguide ports
up to 110 GHz (scalar power calibration only)
A supported coaxial power meter with an upper frequency limit of 23 GHz
or higher for calibrating the LO IN and RF IN signals at the converter (sca-
lar power calibration)
For system error correction, a calibration kit for the respective WM-xxxx wave-
guide
Required equipment