Command reference
R&S
®
ZNA
1054User Manual 1178.6462.02 ─ 20
Table 7-5: Example: frequency-dependent attenuation table
[i] Frequency Attenuation
1 1 GHz 0.01 dB/m
2 2 GHz 0.015 dB/m
3 3 GHz 0.012 dB/m
CALCulate<Chn>:TRANsform:DTFault:DELete <DtfDeleteCable>
Deletes the user-defined cable type with the given name.
Suffix:
<Chn>
.
Channel number
This suffix is ignored: cable types are defined for (and deleted
from) all channels.
Setting parameters:
<DtfDeleteCable> Name of a user-defined cable type.
Example:
CALCulate:TRANsform:DTFault:DELete 'My cable
type'
Deletes the user-defined cable type "My cable type".
Usage: Setting only
Options: R&S ZNA-K2
Manual operation: See "Add / Delete" on page 767
CALCulate<Chn>:TRANsform:DTFault:SELect <DtfSelectCable>
Selects one of the available (predefined or user-defined) cable types for Distance to
Fault measurements.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<DtfSelectCable> Name of the cable type.
Example:
CALCulate1:TRANsform:DTFault:SELect '5088-HFLR'
Selects cable type "5088-HFLR".
Options: R&S ZNA-K2
Manual operation: See "Cable Type..." on page 764
CALCulate<Chn>:TRANsform:DTFault:CENTer <Center>
CALCulate<Chn>:TRANsform:DTFault:SPAN <Span>
Defines the distance window of the Distance to Fault measurement using its center
and span.
SCPI command reference