Command reference
R&S
®
ZNA
894User Manual 1178.6462.02 ─ 20
Example: 
SWE:COUN 10
Define the number of sweeps (10) to be measured in single
sweep mode.
INIT:CONT OFF; :INIT; *OPC?
Activate single sweep mode and start a single sweep sequence
in channel no. 1. Wait until the single sweep sequence is com-
plete.
CALC:DATA:NSW? SDAT,3
Query the results of the 8
th
 sweep.
See also Chapter 8.2.4.3, "Retrieving the results of previous
sweeps", on page 1613.
Usage: 
Query only
CALCulate<Chn>:DATA:NSWeep:COUNt?
Reads the number of completed sweeps in single sweep mode (INITiate<Ch>:
CONTinuous OFF). The trace can be any of the traces acquired during the single
sweep cycle.
Tip:
This command can only be used for [SENSe<Ch>:]SWEep:COUNt > 1.
Suffix: 
<Chn>
.
Channel number used to identify the active trace
Example:  See CALCulate<Chn>:DATA:NSWeep:FIRSt?
Usage:  Query only
CALCulate<Chn>:DATA:NSWeep:FIRSt? <Format>, <FwCount>[, <FwCountEnd>]
Reads the response values of a trace or a consecutive group of traces acquired in sin-
gle sweep mode (INITiate<Ch>:CONTinuous OFF).
Tip:
This command can only be used for [SENSe<Ch>:]SWEep:COUNt > 1.
Suffix: 
<Chn>
.
Channel number used to identify the active trace
Query parameters: 
<Format> SDATa
Read unformatted sweep data (fixed parameter): Returns the
real and imaginary part of each measurement point (2 values
per trace point irrespective of the selected trace format).
SCPI command reference