Setting parameters:
<InputDir> Input Directory
Example:
MMEMory:STORe:CABLe 'C:\dtf-cables'
Saves all predefined and user-defined Distance to Fault (DtF)
cable types to the directory C:\dtf-cables.
Edit *.rsc files for user defined cable types, copy *.rsc files,
rename and edit them. Afterwards execute
MMEMory:LOAD:CABLe 'C:\dtf-cables'
to update existing and create new user-defined cable types.
Usage: Setting only
Options: R&S ZNA-K2
Manual operation: See "Add / Delete" on page 767
MMEMory:LOAD:CKIT <CalKitFile>
Loads cal kit data from the specified cal kit file.
Setting parameters:
<CalKitFile> String parameter to specify name and directory of the loaded cal
kit file. If no path is specified the analyzer searches the current
directory, to be queried with MMEMory:CDIRectory
Example:
MMEM:LOAD:CKIT 'C:\Users\Public\Documents
\Rohde-Schwarz\ZNA\Calibration\Kits\New_kit.calkit'
Load the previously created cal kit file New_kit.calkit from
the default cal kit directory.
... :MMEM:STOR:CKIT 'New_kit', 'C:
\Users\Public\Documents
\Rohde-Schwarz\ZNA\Calibration\Kits\New_kit.calkit'
Store the data for the user-defined cal kit Newkit and overwrite
the cal kit file New_kit.calkit.
Usage: Setting only
Manual operation: See "Import Cal Kit..." on page 535
MMEMory:LOAD:CKIT:SDATa <ConnectorType>, <CalKitName>, <StandardType>,
<StandardLabel>, <TouchstoneFile>[, <FirstPort>[, <SecondPort>]]
Loads cal kit data for a calibration standard from a specified Touchstone file. A restric-
tion on the port assignment may be defined in addition.
Use the newer command MMEMory:LOAD:CKIT:SDATa:WLABel to be able to distin-
guish cal kits by label.
Setting parameters:
<ConnectorType> String parameter containing the name of the connector type.
<CalKitName> String parameter containing the name of a calibration kit availa-
ble on the analyzer.
SCPI command reference