Command reference
R&S
®
ZNA
1214User Manual 1178.6462.02 ─ 20
<DecSeparator> POINt | COMMa
POINt - Decimal separator: point.
COMMa - Decimal separator: comma.
If the third optional parameter is omitted, points are used.
<FieldSeparator> SEMicolon | COMMa | TABulator | SPACe
SEMicolon - Field separator: semicolon
COMMa - field separator: comma.
TABulator - field separator: tabulator.
SPACe - field separator: space.
If the third optional parameter is omitted, semicolons are used.
Example:
*RST; :CONF:TRAC:NAME?
Reset the instrument, creating a default channel no 1 and a
default trace Trc1.
CALC:PAR:DEF:SGR 1,2
Create four traces to measure the two-port S-parameters S
11
,
S
12
, S
21
, S
22
. The traces are not displayed.
MMEM:STOR:TRAC:CHAN 1, 'C:\Users\Public
\Documents\Rohde-Schwarz\ZNA\Traces\Chn1.csv'
Store all trace data of channel 1 to a trace file.
MMEM:STOR:TRAC:CHAN 1, 'C:\Users\Public
\Documents\Rohde-Schwarz\ZNA\Traces\Chn1.s2p'
Store the four S-parameter traces to a two-port Touchstone file.
The Touchstone file will not contain the default trace Trc1.
Usage:
Setting only
Manual operation: See "Save" on page 426
MMEMory:STORe:TRACe:OPTion:BALanced <Boolean>
MMEMory:STORe:TRACe:OPTion:SYMMetric <Boolean>
This command enables/disables the export of balanced (and mixed-mode) S-parame-
ters for MMEMory:STORe:TRACe:PORTs.
Parameters:
<Boolean>
MMEMory:STORe:TRACe:OPTion:COMMent <Comment>
Defines a comment to be added to (the comment section of) exported trace files.
This is a global setting, i.e. the comment string is used for each trace export in every
open recall set.
Parameters:
<Comment> Comment string. If empty, no comment will be added.
Manual operation: See "Comment Added to File" on page 427
SCPI command reference