Command reference
R&S
®
ZNA
1234User Manual 1178.6462.02 ─ 20
[SENSe:]CONVerter:DEFinition:FREQuency <Type>, <StartFreq>, <StopFreq>
[SENSe:]CONVerter:DEFinition:FREQuency? <Type>
Sets/queries the start and stop frequency of (the converter port of) the related con-
verter type.
Parameters:
<StartFreq> Default unit: Hz
<StopFreq> Default unit: Hz
Parameters for setting and query:
<Type> Name of the converter type, whose start and stop frequencies
are set/queried.
Options: R&S ZNA-K8
Manual operation: See "Converter Port: Start Freq./Stop Freq." on page 831
[SENSe:]CONVerter:DEFinition:IFRequency
<Type>, <IF>
[SENSe:]CONVerter:DEFinition:IFRequency? <Type>
Sets/queries the fixed intermediate frequency of the related converter type.
Parameters:
<IF> Intermediate frequency
Default unit: Hz
Parameters for setting and query:
<Type> Name of the converter type, whose IF is set/queried.
Options: R&S ZNA-K8
Manual operation: See "IF" on page 831
[SENSe:]CONVerter:DEFinition:LMUL <Type>, <LocalMultiplicator>
[SENSe:]CONVerter:DEFinition:LMUL? <Type>
Sets/queries the LO multiplicator of the related converter type.
Parameters:
<LocalMultiplicator>
Parameters for setting and query:
<Type> Name of the converter type, whose LO multiplier is set/queried.
Options: R&S ZNA-K8
Manual operation: See "Local Port: Multiplicator" on page 832
[SENSe:]CONVerter:DEFinition:LPOWer <Type>, <LocalNomPower>
[SENSe:]CONVerter:DEFinition:LPOWer? <Type>
Sets/queries the preferred input power at the LO In port of the related converter type.
SCPI command reference