Command reference
R&S
®
ZNA
1253User Manual 1178.6462.02 ─ 20
<L0>, <L1>,
<L2>, <L3>
mandatory* mandatory mandatory* mandatory** mandatory**
OPEN |
SHORt |
MATCh |
<Resist-
ance>
optional*** optional*** optional*** mandatory:
OPEN |
SHORt |
<Resist-
ance>
mandatory:
OPEN |
SHORt |
<Resist-
ance>
<Port1>[,
<Port2>]
optional: <Port 1> optional: <Port 1>,<Port 2>
* values are ignored during calibration
** if OPEN is selected the residual inductance (L0,...,L3) are during calibration; if SHORt is selected the fringing capacitance
(C0,...,C3) are during calibration
*** must be provided if the definition is restricted to a single port. Values are ignored during calibration
The parameters in the [SENSe<Ch>:]CORRection:CKIT:<StandardType>,
[SENSe<Ch>:]CORRection:CKIT:<StandardType>:WLABels, and
[SENSe<Ch>:]CORRection:CKIT:<ConnectorType>:<StandardType> com-
mands have the following meaning:
Table 7-14: Parameter list
Parameter Meaning Comment/Unit
'<ConnType>' Name of the connector type.
Use SENSe1:CORRection:CONNection:CATalog? to query connec-
tor names.
String parameter
'<CalKitName>' Name of the calibration kit. Use
[SENSe:]CORRection:CKIT:CATalog? to query cal kit names.
String parameter
'<CalKitLabel>' Label (e.g. the serial number) of the calibration kit; for ...WLABel com-
mand only
String parameter
'<StandardLa-
bel>'
Label (e.g. the serial number) of the standard String parameter
<MinFreq>,
<MaxFreq>
Min./max. frequency for which the circuit model is valid Default unit is Hz
<DelayParam>
Depending on the cal kit's model type (selected using [SENSe:]CORRection:CKIT:DMODe), this parameter
is interpreted as:
delay [s] for Keysight modeling
el. length [m] for ZVR compatible modeling
Cal kits that are not created/modified using [SENSe:]CORRection:CKIT:DMODe use the ZVR compatible
modeling.
<Loss> Loss (offset parameter) of the standard Must be specified without unit
(implicit unit is dB)
<Z0> Reference impedance (no unit) Must be specified without unit
(implicit unit is Ω)
<C0>,...,<C3> Polynomial coefficients for the fringing capacitance of the standard (load
parameter)
Must be specified without unit:
implicit unit of <Ci> is fF / (GHz)
i
<L0>,...,<L3> Polynomial coefficients for the residual inductance of the standard (load
parameter)
Must be specified without unit:
implicit unit of <Li> is pH / (GHz)
i
SCPI command reference