Command reference
R&S
®
ZNA
1268User Manual 1178.6462.02 ─ 20
Manual operation:  See "Detect Port Assignment" on page 526
[SENSe<Ch>:]CORRection:COLLect:AUTO:TERMination <ReceivingPort>
Acquires receiver noise data in Channel <Ch>, with a matched load connected to
<ReceivingPort>. This is part of a noise figure calibration using a calibration unit..
Suffix: 
<Ch>
.
Channel number
Setting parameters: 
<ReceivingPort> Port number of the receiving port
Example: 
SENSe1:CORRection:COLLect:AUTO:TERMination 2
Acquire noise data at receive port 2.
Usage:  Setting only
Options:  R&S ZNA-K30
[SENSe:]CORRection:COLLect:AVERage 
<Average>
Activates automatic averaging, which means that the VNA may perform multiple cali-
bration sweeps and apply averaging to reduce trace noise. In contrast to manual aver-
aging (see [SENSe<Ch>:]AVERage[:STATe]) the number of calibration sweeps is
calculated automatically.
Parameters:
<Average> AUTO | MANual
Manual operation:  See "Detector" on page 590
[SENSe:]CORRection:COLLect:CHANnels:ALL <Boolean>
Enables calibration of all channels in the active recall set. The command is valid for the
following calibration methods:
●
Manual system error correction
●
Automatic system error correction (cal unit)
●
Manual SMARterCal
●
Automatic SMARTerCal (cal unit)
A scalar power calibration is not affected.
Parameters:
<Boolean> ON – calibrate all channels.
OFF – calibrate the active channel only.
Example:  See [SENSe<Ch>:]CORRection:COLLect:SAVE:
SELected[:DUMMy]
SCPI command reference