Command reference
R&S
®
ZNA
1293User Manual 1178.6462.02 ─ 20
[SENSe<Ch>:]CORRection:COLLect:DELete [<CalName>]
Deletes system error correction data generated and stored previously.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<CalName> Name of the calibration (string parameter) defined together with
the calibration type ([SENSe<Ch>:]CORRection:COLLect:
METHod:DEFine).
ALL - the analyzer deletes all calibrations.
If nothing is specified the analyzer deletes the last system error
correction stored by means of [SENSe<Ch>:]CORRection:
COLLect:SAVE:SELected[:DUMMy].
*RST: -
Example:
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard
as calibration type.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the
measurement results of this standard.
CORR:COLL:SAVE:SEL
CORR:COLL:DEL ALL
Calculate the system error correction data and apply it to the
active channel, then delete the data.
Usage: Setting only
Manual operation: See "Apply" on page 522
[SENSe<Ch>:]CORRection:COLLect:DISCard
Terminates a system error correction, discarding the acquired data.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Example:
:SENSe1:CORRection:COLLect:METHod:DEFine 'Test', TOSM, 1, 2
:SENSe1:CORRection:COLLect:ACQuire:SELected THRough, 1, 2
:SENSe1:CORRection:COLLect:ACQuire:SELected OPEN, 1
:SENSe1:CORRection:COLLect:ACQuire:SELected SHORt, 1
:SENSe1:CORRection:COLLect:ACQuire:SELected MATCh, 1
:SENSe1:CORRection:COLLect:ACQuire:SELected OPEN, 2
:SENSe1:CORRection:COLLect:ACQuire:SELected MATCh, 2
:SENSe1:CORRection:COLLect:DISCard
Usage: Event
SCPI command reference