Command reference
R&S
®
ZNA
1315User Manual 1178.6462.02 ─ 20
<PhyPt> Port number of the analyzer. This numeric suffix is ignored; the
command affects the parameters of all ports.
Parameters:
<Boolean> The parameter function depends on whether the command is
used as a setting command or as a query: For setting command:
ON - no effect
OFF - resets all length offsets to zero and the reference fre-
quency to 1 GHz
For query:
1 - at least one length offset parameter is different from its
default value
0 - all length offsets are zero / set to default
*RST: OFF
Example:
*RST; :CORR:OFFS?
Reset the instrument and query whether the length offset param-
eters have been reset as well. The response is 0.
Manual operation: See "Reset Offsets" on page 653
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:COMPensation[:STATe] <Boolean>
Toggles length/loss/fixture compensation for physical port <PhyPt> ON/OFF.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
Parameters:
<Boolean> ON (1): compensation active
OFF (0): compensation inactive
Manual operation: See "Active" on page 652
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:DFComp[:STATe]?
Returns whether a direct fixture compensation has been carried out at port no.
<PhyPt>. A direct fixture compensation resets the offset parameters to zero, the ana-
lyzer uses calculated transmission factors instead.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt> Port number of the analyzer.
Return values:
<Boolean> 1 - direct fixture compensation data used
0 - no direct fixture compensation data used
Example:
*RST; CORR:OFFS:DFC?
Reset the instrument and query whether the analyzer uses direct
fixture compensation data at port 1. The response is 0.
SCPI command reference