EasyManua.ls Logo

R&S ZNA26 - Page 1383

R&S ZNA26
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1383User Manual 1178.6462.02 ─ 20
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings.
SEGM:DEL:ALL
Delete the created segment and all segments in the channel cre-
ated before.
Usage: Event
Manual operation: See "Add / Insert / Delete / Delete All" on page 500
[SENSe<Ch>:]SEGMent<Seg>:DELete[:DUMMy]
Deletes the specified (single) sweep segment. [SENSe<Ch>:]SEGMent<Seg>:
DELete:ALL deletes all segments in the channel.
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings.
SEGM:DEL
Delete the created segment.
Manual operation: See "Add / Insert / Delete / Delete All" on page 500
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:CENTer?
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:SPAN?
These commands return the center frequency and the span (width) of sweep segment
no. <Seg>.
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz.
SEGM:FREQ:CENT? SPAN?
Query the center frequency and span of the created segment.
The response is 1250000;500000.
Usage: Query only
Manual operation: See "Table Columns" on page 500
Note: The frequency range of the sweep segment can be changed via [SENSe<Ch>:
]SEGMent<Seg>:FREQuency:STARt and [SENSe<Ch>:]SEGMent<Seg>:
FREQuency:STOP.