EasyManua.ls Logo

R&S ZNA26 - Page 1386

R&S ZNA26
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1386User Manual 1178.6462.02 ─ 20
<LO> POSitive | NEGative | AUTO
Position of the local oscillator frequency LO relative to the RF
frequency. In remote control this parameter must be set when a
sweep segment is created.
See [SENSe<Ch>:]FREQuency:SBANd.
<Selectivity> NORMal | MEDium | HIGH
Selectivity of the IF filter. See [SENSe<Ch>:]SEGMent<Seg>:
BWIDth[:RESolution]:SELect.
<Mode> AUTO
Ignored.
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz.
SEGM2:ADD
Create a second sweep segment. The frequency range of the
second segment will be between 1.5 MHz and the maximum fre-
quency of the analyzer.
Manual operation: See "Add / Insert / Delete / Delete All" on page 500
[SENSe<Ch>:]SEGMent<Seg>:INSert:SELect <TimeRef>
Defines whether the sweep time of a new segment, i.e. numeric parameter no. 5 of the
command [SENSe<Ch>:]SEGMent<Seg>:INSert, is entered as a segment sweep
time ("Segment Time") or as a measurement delay ("Meas Delay").
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number
Parameters:
<TimeRef> SWTime | DWELl
SWTime - use segment sweep time.
DWELl - use measurement delay.
*RST: SWTime
Example:
SEGM1:INS:SEL DWEL
Select the meas. delay to determine the sweep time in a new
sweep segment no. 1.
SEGM1:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.01S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz and a meas. delay of 10 ms.
SEGM1:SWE:TIME?
Query the sweep time in the new segment.
Manual operation: See "Optional Columns" on page 503
SCPI command reference

Table of Contents

Other manuals for R&S ZNA26

Related product manuals