Command reference
R&S
®
ZNA
1390User Manual 1178.6462.02 ─ 20
LNOise
10 dB gain, for low input levels
*RST: LDIStortion
Example:
*RST; :SENSe:POWer:GAINcontrol:GLOBal MANual
Enable the manual configuration of independent GC settings for
each drive port.
:SENSe:SEGMent1:ADD; :SENSe:SWEep:TYPE SEGMent
Create a new sweep segment no. 1 and activate segmented
sweep type.
SENSe:SEGMent1:POWer:GAINcontrol:CONTrol ON
Enable segmented GC.
SENSe:SEGMent1:POWer:GAINcontrol:ALL LNO
For sweep segment no. 1, set the GC mode of all a and b wave
receivers to "Low Noise". Segmented GC is enabled automati-
cally.
Usage:
Setting only
Firmware/software: V2.75 or higher
Manual operation: See "Range" on page 636
[SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol:CONTrol <Boolean>
Defines whether common or independent gain control (GC) settings shall be used for
the individual sweep segments.
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number. This suffix is ignored.
Parameters:
<Boolean> ON – use independent GC settings, to be defined via
[SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol.
OFF – reset the GC mode in all sweep segments to the GC
mode for unsegmented sweeps, defined via [SENSe<Ch>:
]POWer:GAINcontrol or [SENSe<Ch>:]POWer:
GAINcontrol:GLOBal. ON will not restore the previous val-
ues.
*RST: ON
Example: See [SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol
Manual operation: See "Segmented IF Gain" on page 634
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]:CONTrol <Boolean>
Selects common or independent internal source "Power" settings for the sweep seg-
ments.
SCPI command reference