Command reference
R&S
®
ZNA
1506User Manual 1178.6462.02 ─ 20
SYSTem:ERRor:COUNt?
Gets the current number of entries in the SCPI error queue.
Usage: Query only
SYSTem:ERRor[:NEXT]?
Queries and at the same time deletes the oldest entry in the error queue. Operation is
identical to that of STATus:QUEue[:NEXT]?
The entry consists of an error number and a short description of the error. Positive
error numbers are instrument-dependent. Negative error numbers are reserved by the
SCPI standard; see section Error messages and troubleshooting.
Example:
SYST:ERR?
Query the oldest entry in the error queue. 0,"No error" is
returned if the error queue is empty.
Usage: Query only
SYSTem:ERRor:ALL?
Queries and at the same time deletes all entries in the error queue.
The entries consist of an error number and a short description of the error. Positive
error numbers are instrument-dependent. Negative error numbers are reserved by the
SCPI standard; see section Error messages and troubleshooting.
Example:
SYST:ERR:ALL?
Query all entries in the error queue. 0,"No error" is returned
if the error queue is empty.
Usage: Query only
SYSTem:ERRor:DISPlay[:REMote] <Boolean>
Disables or enables the display of information popups for remote control errors. These
popups appear at the bottom of the remote screen and the manual screen.
Note
Display of information popups can be globally disabled/enabled using SYSTem:
ERRor:DISPlay:STATe OFF/ON
For SCPI error -113, Undefined header no tooltip is displayed
Parameters:
<Boolean> ON | OFF - enable/disable display of information popups for
remote control errors
*RST: ON