Command reference
R&S
®
ZNA
1516User Manual 1178.6462.02 ─ 20
HALT
Windows is shutdown afer a time-out period of 20 seconds
REBoot
Windows is restarted afer a time-out period of 20 seconds
ABORt
Abort a Windows shutdown/restart. This can only be used during
the time-out period.
CLOSe
Close the firmware.
RESTart
Restart the firmware.
Example:
SYST:SHUT
Switch the analyzer to standby state.
Usage: Setting only
SYSTem:SOUNd:ALARm[:STATe]
<Boolean>
SYSTem:SOUNd:STATus[:STATe] <Boolean>
These commands switch alarm or status sounds on or off.
Parameters:
<Boolean> ON | OFF
*RST: ON
Example:
SYST:SOUN:ALAR OFF; STAT OFF
Switch alarm and status sounds off.
Manual operation: See "Sounds" on page 791
SYSTem:TIME <Hours>, <Minutes>, <Seconds>
The command queries or defines the instrument's current time setting.
The setting command requires administrator rights; refer to "User accounts and pass-
word protection" on page 34.
Parameters:
<Hours> Range: 0...23
<Minutes> Range: 0...59
<Seconds> Range: 0...59
Example:
SYST:TIME?
Response: 12,0,0 - it is precisely 12 pm.
SCPI command reference