EasyManua.ls Logo

R&S ZNA26 - Page 1518

R&S ZNA26
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1518User Manual 1178.6462.02 ─ 20
ON (1)
New traces are created with
CALCulate<Chn>:LIMit:TTLout<Pt>[:STATe] set to ON
(1).
*RST: Not affected by *RST; factory default is OFF
Manual operation: See "TTL Pass Default Values" on page 797
SYSTem:USER:DISPlay:TITLe <String>
Defines a title for the remote display, i.e. for the screen that is shown at the instrument,
if a remote session is established and the standard display is switched off.
Parameters:
<String> Title string.
A \n in the string starts a new line; see example.
*RST: empty string
Example:
SYST:USER:DISP:TITL "DON'T TOUCH\nRemote test
running..."
Define a title for the remote display.
SYSTem:USER:KEY <Key>[, <Label>]
Labels a user-defined key in the remote display. In the query form the command
returns whether or not a user-defined key was tapped or clicked.
Parameters:
<Key> Number of the user key
0 Delete all user keys and restore the default keys ("Go to
Local", "Display Off").
1 to 8 User key numbers
Range: 0 to 8
*RST: 0
<Label> Label for user key no. 1 to 8 (string variable)
SCPI command reference

Table of Contents

Other manuals for R&S ZNA26

Related product manuals