Example: //Return installed software information
DIAG:HUMS:SW?
Usage: Query only
DIAGnostic:HUMS:SYSTem:INFO?
This command returns the system information of the instrument. Depending on the set
data format, the queried data is either displayed in XML or JSON format. For more infor-
mation about setting the data format, see DIAGnostic:HUMS:FORMat on page 1551.
Return values:
<SystemInfo> <block_data>
Returns system information of the connected instrument as
block data.
Binary block data with the following syntax:
#<Digits><Length><Binarydata>
#
Indicates the start of the binary block
<Digits>
Decimal value
Gives the number of decimal digits used for the <Length>
value
<Length>
Decimal value
Number of bytes the follow in the <Binary data> part
<Binary data>
Binary data in ASCII format
Example:
//Return system information
DIAG:HUMS:SYST:INFO?
Usage: Query only
DIAGnostic:HUMS:SYSTem:STATus?
This command returns the complete system status information of the instrument.
Depending on the set data format, the queried data is either displayed in XML or JSON
format. For more information about setting the data format, see DIAGnostic:HUMS:
FORMat on page 1551.
Return values:
<SystemStatus> <block_data>
Returns complete system status information of the connected
instrument as block data in a comma-separated list:
#blockdata{<globalStatus>,<tablevalues>:
[{tablevalue1},{tablevalue2},...]}