EasyManua.ls Logo

R&S ZNA26 - Page 164

R&S ZNA26
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Concepts and features
R&S
®
ZNA
164User Manual 1178.6462.02 ─ 20
A reflection OSM calibration is more accurate than a normalization but is only applica-
ble for reflection measurements.
4.5.1.3 One Path Two Ports Calibration
A one path two ports calibration combines a reflection OSM (full one-port) calibration
with a transmission normalization. The fully calibrated port is termed the node port.
This calibration type requires a Short, an Open and a Match standard to be connected
to a single test port plus a Through standard between this calibrated source port and
the other load ports. The four standard measurements are used to derive the following
error terms:
The Short and Open standards are used to derive the source match and the reflec-
tion tracking error terms at the source port.
The Match standard is used to derive the directivity error at the source port.
The Through standard provides the transmission tracking error terms.
For calibration of two ports, a one-path two-port calibration requires only four standards
to be connected (instead of 7 for a full two-port TOSM calibration). It is suitable when
only the forward (e.g. S
11
and S
21
) or reverse S-parameters (e.g. S
22
and S
12
) are nee-
ded, and if the DUT is well matched, especially at the load port.
A one-path two-port calibration is also the best calibration method for test setups with
unidirectional signal flow, e.g. a pulsed measurement using an external generator.
4.5.1.4 TOSM and UOSM calibration
TOSM
A TOSM (Through – Open – Short – Match) calibration requires the same standards as
the one path two ports calibration, however, all measurements are performed in the for-
ward and reverse direction. TOSM is also referred to as SOLT (Short – Open – Load =
Match – Through) calibration. The four standards are used to derive 6 error terms for
each signal direction:
In addition to the source match and reflection tracking error terms provided by the
one-path two-port calibration, TOSM also provides the load match.
The directivity error is determined at all source ports.
The transmission tracking is determined for each direction.
TOSM calibration is provided for arbitrary n-port measurements (n>1). The number of
required standard measurements and of error terms increases as shown in the follow-
ing table.
Calibration

Table of Contents

Other manuals for R&S ZNA26

Related product manuals