Concepts and features
R&S
®
ZNA
184User Manual 1178.6462.02 ─ 20
thermal vacuum chamber (TVAC), a recalibration using traditional calibration kits or
calibration units is a costly and time consuming process.
The Rohde & Schwarz inline calibration system (ICS) R&S ZN-Z3x allows recalibrating
a running measurement setup without mechanically changing it. For each DUT port, an
inline calibration unit (ICU) is inserted between the corresponding VNA test port and
the DUT. The ICU remains within the signal path throughout the measurement (low-
loss through path) and, on request of the VNA, performs a full one-port calibration
(OSM) in-situ.
For n-port measurements with n>2, an initial n-port calibration of the test setup has to
be done.
Figure 4-9: ICU R&S
ZN-Z32
Control connections are established via a central Inline Calibration Controler (ICC) with
2 CAN bus interfaces for connecting the ICUs
USB and LAN interface for remote control (from the R&S ZNA or a standard Win-
dows PC)
Calibration