Concepts and features
R&S
®
ZNA
252User Manual 1178.6462.02 ─ 20
period, pulse rise and fall times, power drop across a pulse, and intrapulse phase mod-
ulation, and produces a trend analysis over many pulses.
The R&S ZNA can push pulse profile measurement data to a running R&S VSE
instance for further (live!) analysis.
Requirements
R&S ZNA:
FW V2.20 or higher
with option R&S ZNA-K7 "Pulsed measurements"
R&S VSE:
basic or enterprise edition (licensed separately)
running on the R&S ZNA or on a PC that is reachable via LAN
V1.90 or higher
with option R&S VSE-K6 "Pulse measurements application"
Limitations
only b-waves can be pushed to the R&S VSE
only a single b-wave can be pushed to the R&S VSE at a time
the sweep duration must be 50 µs or higher
Figure 4-25: R&S
VSE with pulse measurement application (option R&S VSE-K6)
How to configure
1. On the target R&S VSE, prepare the required analysis windows