GUI reference
R&S
®
ZNA
327User Manual 1178.6462.02 ─ 20
Remote command:
see "Bandwidth" on page 485
Noise Det Meas Time
For measuring the noise power, the root-mean-square (RMS) of each measurement
sample is calculated. The noise detector measurement time sets the detector time at
each measurement point for the RMS detector, allowing to choose how many RMS val-
ues are used at each frequency. A longer detector time and therefore a higher number
of RMS values results in a closer estimation of the real noise power.
"Auto" means that the value is set from the NF quickset dialog. Use the checkbox to
the right to enable the input field and set its value manually.
Remote command:
[SENSe<Ch>:]NFIGure:NTIMe
Gain Det Meas Time
Sets the detector time at each measurement point for the AVG detector, which is used
for the measurement of the DUT gain. Using the AVG detector allows you to remove
statistical fluctuations from the measurement and therefore achieves a more accurate
result.
"Auto" means that the value is set from the NF quickset dialog. Use the checkbox to
the right to enable the input field and set its value manually.
Remote command:
[SENSe<Ch>:]NFIGure:GTIMe
Cal & Meas Overview
Opens the "Cal & Meas Overview" dialog, which displays all settings related to NF cali-
bration and measurement.
Meas softtool