GUI reference
R&S
®
ZNA
505User Manual 1178.6462.02 ─ 20
Any trigger event can start an entire sweep or a part of it. Moreover, it is possible to
switch off the RF source between consecutive triggered measurement sequences, and
to define a delay between trigger events and the measurement sequences.
Background information
Refer to Chapter 4.1.4.1, "Partial measurements and driving mode", on page 88.
Output trigger
If the R&S ZNA is equipped with the Trigger Board option R&S ZNA-K91, it provides
four configurable output trigger signals that can be used to synchronize external devi-
ces with the measurement. These signals are available at the rear panel connectors
Trigger Out A to D.
5.10.3.1 Controls on the Trigger tab
The trigger settings are also valid for calibration sweeps. Hence, in external trigger
mode, the external trigger signal must be available during the system error correction,
too. To start the calibration sweeps without delay, use the "FreeRun" trigger type.
The following buttons in the "Trigger" tab open related dialogs:
"Trigger Manager...": Trigger Manager dialog
"Trigger Out Manager...": Trigger Out Manager dialog
Sweep Softtool