Loading...
GUI reference
R&S
®
ZNA
514User Manual 1178.6462.02 ─ 20
Related information
Refer to Chapter 4.1.4, "Sweep control", on page 87.
Sweep All Channels/Sweeps
If you select "Sweep All Channels", Restart Sweep restarts the sweep in all channels
of the active recall set.
For the DEFAULT Remote Language, the number of sweeps in single sweep mode is
channel-specific and can be defined in the table below the separator line. The
"Sweeps" column reflects the Sweeps settings of the individual channels. The
"Sweeps" setting above the separator line is unused and disabled.
For remote languages other than DEFAULT, the "Sweeps" setting above the separator
line defines the common number of sweeps that are executed on each channel in sin-
gle sweep mode. The table below the separator line is unused and disabled.
Tip: In remote control, it is possible to retrieve the results acquired in any of the
sweeps within a single sweep group.
Remote command:
INITiate[:IMMediate]:ALL
INITiate<Ch>[:IMMediate]:SCOPe ALL
[SENSe<Ch>:]SWEep:COUNt
Sweep Active Channel/Sweeps (table)
If you select "Sweep Active Channel", then in single sweep mode Restart Sweep only
restarts the sweep in the active channel.
The "Sweeps" column reflects the Sweeps settings of the individual channels.
Remote command:
INITiate<Ch>[:IMMediate]:SCOPe SINGle
[SENSe<Ch>:]SWEep:COUNt
Sweep Softtool