GUI reference
R&S
®
ZNA
538User Manual 1178.6462.02 ─ 20
Cal. Power Settings tab
The "Cal Power Setting" tab collects all calibration-related power settlings that were
previously distributed across several softtool tabs and dialogs.
Source Power
See "Power" on page 477.
Source Step Att
Source step attenuator setting used during calibration, if activated.
Remote command:
[SENSe<Ch>:]CORRection:ADVanced:SOURce<PhyPt>:ATTenuator
on page 1242
[SENSe<Ch>:]CORRection:ADVanced:SOURce<PhyPt>:ATTenuator:STATe
on page 1243
ZNA Port Power
"Power Result", as displayed in the port settings dialog.
Cal Power Offset
For power calibrations only: Specifies a gain (positive values) or an attenuation (nega-
tive values) in the signal path between the source port and the calibrated reference
plane. With a "Cal Power Offset" of n dB, the target power at the reference plane (cal
power) is equal to the actual output power at the port plus n dB. The "Cal Power Off-
set" has no impact on the source power.
Cal softtool