GUI reference
R&S
®
ZNA
543User Manual 1178.6462.02 ─ 20
During a measurement receiver calibration ("Meas. Receiver"), the following traces are
displayed:
The trace "a<m>(P<m>)" shows the (previously calibrated) power at the calibrated
reference plane (source port P<m>).
The trace "b<n>(P<m>)" shows the current power reading of the analyzer at the
calibrated receive port P<n> (source port P<m>).
After successful measurement receiver calibration, the "b<n>(P<m>)" trace must be
close to the "a<m>(P<m>)" trace. Due to the previous power calibration, both traces
must be close to the cal power.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]:VERification:
RESult?
Port Overview
Shows all source ports together with the possible power calibrations.
For each analyzer port P1 ... PN, a "Ref. Receiver", a "Meas. Receiver" and a "Source
Flatness" calibration can be performed.
If a "SMARTerCal" is active, additional receiver power calibrations are not allowed
and the R&S ZNA only offers the possibility to perform "Source Flatness" calibra-
tions of the related test ports:
See Chapter 4.5.7.3, "Combining SMARTerCal with scalar power calibration",
on page 199 for details.
If external generators are configured, they appear as additional source ports G1 ...
Gm in the port overview. A source power (flatness) calibration is available at each
generator port.
If the optional LO Out is equipped and not configured as LO source for millimeter
wave converters, it appears as an additional "C.LO" port.
Only a source flatness calibration can be performed, and this calibration can only
be done using a power meter (irrespective of the selected Power Cal Method.
If option R&S ZNA is available and frequency converters are configured, additional
source flatness calibrations can be performed for the analyzer ports driving the
converters' RF IN and LO IN ports.
Cal softtool

Table of Contents