GUI reference
R&S
®
ZNA
545User Manual 1178.6462.02 ─ 20
The power calibration state is indicated in the trace list, see Chapter 4.5.6.3, "Power
calibration labels", on page 195. Use the functions in the Chapter 5.11.4, "Use Cal tab",
on page 579 to activate, deactivate, or store power calibrations.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
SOURce<Ch>:POWer:CORRection:DATA
SOURce<Ch>:POWer:CORRection:DATA:PORT<PhyPt>
[SENSe<Chn>:]CORRection:PSTate?
Power Cal dialog – Ref. Receiver
In Ref. Receiver mode, the "Power Cal" dialog guides you through a reference receiver
calibration.
The selected source port is displayed with its cal power settings (see Chapter 5.11.3.2,
"Cal Power Config dialog", on page 574). A circuit diagram visualizes the connectivity.
A power meter must be connected to the calibrated source port.
Access: Power Cal wizard > Ref. Receiver
Background information
Refer to Chapter 4.5.6.1, "Source power calibration", on page 193.
Cal softtool