Test Port Assignment
Defines the assignment between test ports and cal unit ports. In the default "Manual"
assignment, VNA ports and cal unit port numbers match. If you decide to use a differ-
ent assignment, you can auto-detect the actual assignment ("Automatic") or select the
analyzer port numbers manually. Auto-detection can fail, e.g., because of a high
attenuation in the signal path.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
[SENSe:]CORRection:COLLect:AUTO:PORTs:CONNection?
[SENSe:]CORRection:COLLect:AUTO:CKIT:PORTs
Take OSM / Take All OSM and Through
"Take OSM" starts a calibration sweep for the related port. "Take All OSM and
Through" initiates a series of calibration sweeps; the R&S ZNA acquires a full set of
one-port and two-port data. The latter is required for the transmission normalizations
and for a "One Path Two Ports" calibration; see "Dependency between calibration
types and characterization data" on page 183.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
[SENSe:]CORRection:COLLect:AUTO:PORTs:CONNection?