The "Uncertainty" checkbox indicates whether the selected characterization actually
comprises uncertainties.
Remote command:
[SENSe:]CORRection:CKIT:SELect
Same Connector All Ports / Same Gender All Ports
Assigns the same connector type or gender to all selected physical ports. For some
multi-port calibration types, the port connector types must be equal, e.g. because they
require a Through standard with known characteristics.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:CONNection:PORTs
[SENSe<Ch>:]CORRection:COLLect:CONNection:GENDers
Import Cal Kit...
Opens the "Import Calibration Standards with Uncertainties" dialog, that allows you to
import cal kits with (or without) uncertainties.
A "CalKit with Uncertainties" can be one of the following:
a zip file containing calibration standard characterizations in a format supported
by METAS VNA Tools (sdatb, sdatx, sdatcv, calstd)
a scolcv file
See the document "Data Formats" on the METAS VNA Tools internet page (https://
www.metas.ch/metas/en/home/fabe/hochfrequenz/vna-tools.html, section "VNA Tools
Documentation".
The imported cal kit is assigned to the selected Connector type; the name of the ZIP
file is used as the cal kit name.
The name of a characterization file inside the zip container must indicate
the standard type (Open, Short, Match, Load, Line, Reflect, Through)
and, if applicable, its gender (f, m, ff, fm, mf, mm).
Imported cal kits with uncertainties are preserved and added to the Cal Kit selection
combo-box. There and in the Calibration Kits dialog, they appear with an "Uncertainty"
flag.