GUI reference
R&S
®
ZNA
740User Manual 1178.6462.02 ─ 20
Remote command:
MMEMory:STORe:EYE:MEASurements
5.17.2.5 Eye Diagram dialog
Allows basic configuration of the eye diagram simulation.
Access: System – [Applic] > "TDR" > "Eye Diag" > "Eye Diagram..."
For advanced configuration of the eye diagram simulation see Chapter 5.17.2.6,
"Advanced Settings dialog", on page 743
Measurement / Topology
"Measurement" allows you to select an S-Parameter, i.e. the transmission (or reflec-
tion) whose time domain properties shall be analyzed. The "Topology" to the right is
updated accordingly.
Remote command:
CALCulate<Ch>:PARameter:MEASure
Bit Stream
Describes the bit stream generated by the "virtual" generator.
"Bit Stream" can either be a pseudo-random binary sequence (PRBS) of the selected
Length or a user defined bit stream.
The PRBSs are generated using a linear feedback shift register with generator polyno-
mials allowing the maximal run length of the sequence.
Remote command:
CALCulate<Chn>:EYE:INPut:BPATtern:TYPE