Command reference
R&S
®
ZNA
953User Manual 1178.6462.02 ─ 20
The resulting Touchstone files are written to C:\Users\Public\Documents\Rohde-
Schwarz\ZNA\Embedding. If result files with the same name already exist, they will be
overwritten.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Example: See Chapter 8.2.8, "Fixture modeling", on page 1628
Usage: Event
Manual operation: See "Run <Fixture Modeling Tool>" on page 679
CALCulate:FMODel:ISD<Ph_pt>:RUN[:STATe]
<Boolean>
For a fixture modeling with the ISD tool, this command allows to specify the ports to
which the deembedding files shall be assigned after the tool has run.
Suffix:
<Ph_pt>
.
Physical port number for single-ended deembedding, logical port
number for balanced deembedding.
Parameters:
<Boolean> 1 (ON) if a deembedding file shall be assinged to port <Pt>, 0
(OFF) otherwise
Example: See Chapter 8.2.8, "Fixture modeling", on page 1628
Manual operation: See "Apply" on page 680
CALCulate:FMODel:ISD<Ph_pt>:SCALe:ATTenuation <AttenLeadInScaling>
Scales the test coupon’s attenuation (dB).
Sets/gets the atten_scale 1 batch mode parameter of the ISD tool
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<AttenLeadInScaling>
Manual operation: See "Scaling for Lead In Atten" on page 682
CALCulate:FMODel:ISD<Ph_pt>:SCALe:FREQuency <MaxFrequency>
Defines the maximum frequency to deembed.
Corresponds to the max_frequency batch mode parameter of the ISD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<MaxFrequency> Default unit: Hz