Preparing for use
R&S
®
ZNA
16Getting Started 1178.6456.02 ─ 18
Regarding length and quality, the following requirements have to be met for cable that
are directly connected to the R&S ZNA:
Table 3-1: Cable Requirements
Cable Type (Connector) Requirement
RF cables (PORT 1, ..., PORT N) Double shielded
BNC cables (various) Double shielded
DB-25 (User Port) Double shielded
Digital I/Q (External Handler IO, External Data Log-
ger, Direct Control)
R&S order no. 1402.4990.00 only
GPIB Standard cable
RFFE/GPIO R&S ZN-Z25 (order no. 1334.3424.02) only
DisplayPort (Monitor) Standard cable
DVI-D (Monitor) 2 ferrite cores
LAN At least CAT6, S/FTP
PCIe Standard cable
USB Standard cables, length ≤ 3m
Signal input and output levels
Information on signal levels is provided in the data sheet. Keep the signal levels within
the specified ranges to avoid damage to the product and connected devices.
Preventing electrostatic discharge (ESD)
Electrostatic discharge is most likely to occur when you connect or disconnect a DUT.
NOTICE! Risk of electrostatic discharge. Electrostatic discharge can damage the
electronic components of the product and the device under test (DUT).
Ground yourself to prevent electrostatic discharge damage:
a) Use a wrist strap and cord to connect yourself to ground.
b) Use a conductive floor mat and heel strap combination.
During operation, if the firmware observes a serious unexpected disturbance (e.g. due
to ESD), it resets all hardware components to ensure proper instrument functioning. It
then restores the user settings to the state before the disturbance and indicates the
foregone hardware reset by an "Hardware communication problem [...]" information
popup.
3.6 Connecting the analyzer to the AC supply
For safety information, see "Connecting to power" on page 8.
Connecting the analyzer to the AC supply