Notice. Sensor Network addresses must necessarily be different (the program «Config900» it
does not control). Network addresses are chosen from the range 0 ... 5.
Save Configuration button. If necessary, you can also save the configuration to a file button.
The calibration procedure FLS "EPSILON"
Run the program "eS Install" (see details in Appendix 1)
On the tab "General Options sensor", set the following parameters of the measuring head:
• "The network address of the device" from 0 to 5, is unique for each sensor
• "Repeat Interval": it does not matter
• "Bit" 10-bit
• "On. Periodically issue": disabled
• "On. Text format": disabled
• "The rate of exchange on the UART": 19200 bits/s