Pantone 534 Blue
Pantone 123 Yellow
Pantone 485 Red
Pantone 123 Yellow
Pantone 534 Blue
Black
Rich Black -
20/20/20/100
Blue - 100/80/30/5
Yellow - 0/27/100/0
Red - 10/100/100/5
Yellow - 0/27/100/0
Blue - 100/80/30/5
Only if you REALLY need them:
Pantone 534 Blue - 100/80/30/5
Pantone 485 Red - 10/100/100/5
Pantone 123 Yellow - 0/27/100/
AC Current
50 to 1000Hz
400μA 0.1μA
±(1.5% reading + 3 digits)
4000μA 1μA
40mA 0.01mA
400mA 0.1mA
10A 0.01A
(20A: 30 sec max with reduced accuracy)
All AC voltage ranges are specied from 5% of range to
100% of range
NOTE: Accuracy is stated at 65
o
F to 83
o
F (18
o
C to 28
o
C) & less than 75%RH
Function Range Resolution Accuracy
Resistance
400Ω 0.01Ω ±(0.3% reading + 9 digits)
4kΩ 0.0001kΩ
±(0.3% reading + 4 digits)
40kΩ 0.001kΩ
400kΩ 0.01kΩ
4MΩ 0.001MΩ
40MΩ 0.001MΩ ±(2.0% reading + 10 digits)
Capacitance
40nF 0.001nF
±(3.5% reading + 40 digits)
400nF 0.01nF
4μF 0.0001μF
±(3.5% reading + 10 digits)40μF 0.001μF
400μF 0.01μF
4000µF 0.1µF
±(5% reading + 10 digits)
40mF 0.001mF
Frequency
(electronic)
40Hz 0.001Hz
±(0.1% reading + 1 digits)
400Hz 0.01Hz
4kHz 0.0001kHz
40kHz 0.001kHz
400kHz 0.01kHz
4MHz 0.0001MHz
40MHz 0.001MHz
100MHz 0.01MHz Not specied
Sensitivity: 0.8V rms min. @ 20% to 80% duty cycle and
<100kHz; 5Vrms min @ 20% to 80% duty cycle and > 100kHz.
Frequency
(electrical)
40.00-
400Hz
0.01Hz ±(0.5% reading)
Sensitivity: 15Vrms
Duty Cycle
0.1 to
99.90%
0.01% ±(1.2% reading + 2 digits)
Pulse width: 100µs - 100ms, Frequency: 5Hz to 150kHz
Temp (type-K)
-58 to
1832°F
1°F
±(1.0% reading + 4.5°F)
±(1.0% reading + 2.5°C)
(probe accuracy not included)
-50 to
1000°C
1°C
4-20mA%
-25 to
125%
0.01% ±50 digits
0mA=-25%, 4mA=0%, 20mA=100%, 24mA=125%
Note: Accuracy specications consist of two elements:
• (% reading) – This is the accuracy of the measurement circuit
continued ...