9.1 Overload display
If the weight applied to the platform exceeds the maximum
capacity of the REFSCALE-S (110 kg / 243 lb), the display will show
"OOR", which means "Out of Range". This applies to the absolute
weight measurement.
To avoid damage to the REFSCALE-S, remove the weight
from the platform immediately.
9.2 Troubleshooting Guide
Device cannot be
switched on
Battery inserted incorrectly.
Battery contact break was not pulled
out.
Weight over 110 kg.
Reduce weight immediately.
Stable measured
value is not met
Re-align the device.
Restart the device.
Hold function is activated (Hold function
symbol is displayed).
Bluetooth
connection doesn't
work
A device is already connected. This can
be seen when the Wireless connection
icon lights up continuously.
Exceeding the maximum permitted
deviation limit of 125 g