9.0 MENU OPTIONS
A Exit Menu – Pressing the MENU key when this is
displayed will exit the menu mode.
B Calibration – Displays options of:
1. Set (water) - Pressing MENU will prompt to Apply
Water and then “CAL” or “MENU” key to cancel.
2. Span 1.40229 – Press MENU will prompt to
Apply 1.40229, NIST certified calibration oil and
to press “CAL” or “MENU” key to cancel.
3. Span 1.51416 – Press MENU will prompt to
Apply 1.51416, NIST certified calibration oil and
to press “CAL” or “MENU” key to cancel.
4. Add/Delete/Edit – When selecting any of these
modes a “Warning” will appear “Incorrect use of
this function can corrupt the instrument”. Contact
the Technical Support Group for a complete
explanation if this function is to be used.
5. Exit - Press MENU to exit the calibration options
mode.
C Contrast – Press the MENU key to enter and the Up/
Down keys to adjust display contrast.
D IR Linkup – To use this function the optional Docking
Station (13951000) must be used. This feature
allows for downloading of stored values, up to 225 of
the most current readings and up to 5 point current
calibration log.
E Set Clock – Allows for setting of the time in 12 hour
am/pm or 24 hour military and the date in MDY
format.
F Select Channel – This feature is visible only when at
least one custom channel has been added to the
AR200. Entering this feature allows the operator to
switch between available channels/scales.
G Reading Mode – Allows for changing between nD,
nD-TC, Brix or Brix-TC reading modes. TC modes
are Temperature Compensated based upon Sucrose
data in channel 1.