English 21
Pressure accuracy
Maximum static pressure variation at 10 cm H
2
O (10 hPa) according to
ISO 80601-2-70:2015
2
2
2
2
Maximum dynamic pressure variation according to ISO 80601-2-70:2015
Device without humidification and Standard air tubing / Device with humidification and Standard air tubing
2
25 0.3 / 0.3 0.5 / 0.4 0.7 / 0.7
Device without humidification and SlimLine air tubing / Device with humidification and SlimLine air tubing
2
8 0.5 / 0.5 0.5 / 0.5 0.8 / 0.8
20 0.5 / 0.5 0.5 / 0.5 0.8 / 0.8
Pressure accuracy - bilevel
Maximum dynamic pressure variation according to ISO 80601-2-70:2015.
Device without humidification and Standard air tubing / Device with humidification and Standard air tubing
rate
Inspiratory pressure (cm H
2
O [hPa]) (Means, Standard Deviations)
-0.09, 0.01 / -0.22,
0.01
-0.01, 0.07 / -0.22,
0.01
-0.03, 0.09 / -0.29,
0.03
0.02, 0.08 / -0.22,
0.01
0.12, 0.01 / -0.22,
0.01
0.15, 0.01 / -0.26,
0.01
0.15, 0.01 / -0.31,
0.02
0.16, 0.12 / -0.30,
0.02
0.01
0.01
0.01
0.02
0.03
rate
Expiratory pressure (cm H
2
O [hPa]) (Means, Standard Deviations)
-0.14, 0.01 / -0.27,
0.01
-0.16, 0.01 / -0.29,
0.02
-0.11, 0.10 / -0.34,
0.02
-0.16, 0.05 / -0.33,
0.01
-0.17, 0.05 / -0.33,
0.02
0.01
0.02
0.01
0.02
0.02
-0.27, 0.01 / -0.37,
0.01
-0.26, 0.02 / -0.34,
0.01
-0.25, 0.01 / -0.38,
0.01
-0.29, 0.01 / -0.43,
0.02
-0.31, 0.01 / -0.45,
0.03