• To add supplemental oxygen:
1. Unlock the low flow oxygen inlet at the rear of the device by pushing up on the locking clip
2. Insert one end of the oxygen supply tubing into the oxygen port connector. The tubing will
automatically lock into place
3. Attach the other end of the oxygen supply tubing to the oxygen supply
4. Start ventilation
5. Turn on oxygen and adjust (at the oxygen supply) to the prescribed flow rate or FiO2 level
6. Turn off oxygen when you stop ventilation