20
Settings menu
You set all parameters relating to a patient's therapy and device configuration in the Settings menu.
Note: Not all parameters are available in all regions. The default and range values may differ
between modes and regions.
The range of parameters in the Settings menu are expressed in cm H
2
O, where 1cm H
2
O is equal to
0.98 hPa. The units can be changed under Configuration.
Therapy
Mode
Sets the therapy mode
available on the device.
Set Pressure Sets the fixed treatment
pressure.
4–20 cm H
2
O (4–20 hPa),
0.2 cm H
2
O (0.2 hPa)
increments
IPAP Sets the pressure to be
delivered to the patient
when the device is
triggered into inspiration.
4–30 cm H
2
O (4–30 hPa),
0.2 cm H
2
O (0.2 hPa)
increments
EPAP Sets the pressure to be
delivered to the patient
when the device is
cycled into expiration.
2–[IPAP] cm H
2
O (2–
[IPAP] hPa), 0.2 cm H
2
O
(0.2 hPa) increments
Min PS Sets the minimum
pressure support
delivered by the device.
0–20 cm H
2
O (0–20 hPa),
0.2 cm H
2
O (0.2 hPa)
increments
Max PS Sets the maximum
pressure support in
iVAPS mode.
0–28 cm H
2
O (0–28 hPa),
0.2 cm H
2
O (0.2 hPa)
increments
8-28 cm H
2
O (8–28 hPa) when
AutoEPAP enabled.
Min EPAP Sets the minimum EPAP
(minimum expiratory
pressure) delivered by
the device.
Available when
AutoEPAP is enabled.
2–25 cm H
2
O (2–25 hPa),
0.2 cm H
2
O (0.2 hPa)
increments
Max EPAP Sets the maximum EPAP
delivered by the device.
Dependent on Min EPAP.
Available when
AutoEPAP is enabled.
2–25 cm H
2
O (2–25 hPa),
0.2 cm H
2
O (0.2 hPa)
increments
AutoEPAP Enable / disable
AutoEPAP.
Off / On