46
Messages that may show up
Maintenance and cleaning
Clean the outer shell of the device using a clean, damp cloth (do not use aggressive detergent
solutions). The penetration of liquids must be strictly avoided.
During testing operations, headset pads and probe tips come into direct contact with the patient.
Although said components are biocompatible, they are not sterile and they must therefore be
subject to special care:
− After each test session it is necessary to disinfect the headset pads before using them to test
a new patient. To do so, use only commonly marketed non-allergenic disinfectant solutions and
follow the directions supplied by the solution manufacturer.
The transducers and insertion headsets tips (IP30 and Insert) are disposable: use only once
and substitute them before use with a new patient.
Calibration and any possible repair, installation and/or update of the device must be strictly
carried out only by specialized technical personnel, duly authorized by the manufacturer and in
full compliance with the terms and conditions specied in the Technical Manual.
Message on
display
When it appears Message
type
Meaning Troubleshooting
DO YOU
CONFIRM ?
This query appears after entry of
plot cancellation or parameter
change requests
System
message
It means that a plot
cancellation or a para-
meter change request
has been entered
Press either the YES or the NO
when the query appears
CHANGE
CALIBRATION
DATE ?
This message appears at the end
of the device calibration proce-
dure, once the “SAVE” softkey has
been pressed and values have
been changed.
System
message
It provides the option
of updating periodic
calibration dates
Press either the YES or the NO
when the query appears
ENTER
PASSWORD
This message appears when
attempting rst-time access to
the device calibration procedure,
once the unit has been switched
on
System
message
It instructs you to enter
your access password
to unlock the device
calibration procedure
Enter your password via the
softkeys
WRONG
PASSWORD
This message appears when
attempting to access the device
calibration procedure using a
wrong password
System
message
It means that you have
entered the wrong
password for access to
the Calibration menus
Press the Enter key, then re-
enter the password correctly
PARAMETERS
CHANGED
This message is displayed upon
startup of the device in the event
that the system detects possible
errors in one or more than one
parameter relating to the test.
The incorrect parameters will be
displayed simultaneously with
the message, on the same page
System
message
It means that one or
more than one of the
les relating to the test
parameters is either
corrupt or missing
Press the Enter key. The
corrupt les will be replaced
with the system default les
and it will then be possible to
proceed using the device. If
you are instructed to do so, it
will be necessary to run the
device calibration procedure