24
25
Technical Specications
Name of product Revitive Essential
Model RLV
Frequency 20Hz – 53Hz
Output current
Max 8.5mA
Weight
1.13kg
Dimensions Ø 355 mm x 77(D) mm
Power consumption 5W
AC adaptor
CE Approved
Power source
Input (adaptor used)
Output
100-240V AC ( ), 50/60Hz, 0.18A
5V ( ) DC ,1.0A
Service Life of Device
Period of time specified by the manufacturer during
which the device is expected to remain safe for use.
4 Years
Applied parts
Part of Revitive that in normal use come into physical
contact with user to perform therapy
Foot-pads - 135.5cm
2
Electrode Body Pads - 5cm x 5cm = 25cm
2
The values of PULSE DURATIONS, PULSE repetition frequencies and amplitudes, including
any d.c. component, shall not deviate by more than ± 20 % when measured with a load
resistance within the range specified.
If confirmation is required that the Revitive works within its Essential Performance aer a
certain period of time, contact the manufacturer
EN
Output Specifications for Electrical Muscle Stimulation (EMS):
Waveform
Monophasic
Shape
Square symmetrical with polarity reversal
Maximum Output Voltage (+/-15%): @500kΩ Foot: 21.5Vp Body: 21.5Vp
@2kΩ Foot: 55Vp Body: 62.5Vp
@10kΩ Foot: 65Vp Body: 85Vp
Maximum Output Current (+/-15%): @50kΩ Foot: 43mA Body: 43mA
@2kΩ Foot: 27.5mA Body: 31.3mA
@10kΩ Foot: 6.5mA Body: 8.5mA
Pulse Duration
(+/-10%)
:
Foot: 450 or 970µs (Mode 7) Body: 450µs
Frequency (+/-10%): Foot: 20 to 44Hz Body: 37 to 51Hz
Net Charge @ 500Ω
[0.001]mC
Maximum Power Density @ 500Ω
Foot: 0.68 mW/cm² Body: 1.66 mW/cm²
ON Time
Foot: 1.9 - 8.3s Body: 1.9 - 6.3s
OFF Time
Foot: 1.00 - 1.50s
Body: 1.00s
5827_IFU04_17963935.indd 24-25
5827_IFU04_17963935.indd 24-25
09/04/2021 13:23