Problem Possible Cause Solution
Revitive is on (lights
illuminated on the LED
display) but I cannot feel
the electrical stimulation
through the foot-pads.
Not placing both bare feet on the
foot-pads at the same time.
Ensure that your feet are bare and each foot
is placed on each of the foot-pads at the same
time. Keep increasing the intensity up to a
maximum of 50 until you feel the stimulation.
Your feet may be dry. Moisturise the soles of your feet to improve
conductivity and stimulation and try the
procedure again. You may also have to increase
the intensity level.
You may be dehydrated. Drink plenty of water before and aer using the
device. The device uses your body to create
the electrical circuit. Water is an excellent
conductor of electricity and if your body is less
hydrated (below 60%) then the stimulation
may be less, therefore it is important to always
remain well hydrated.
The intensity level may be on too
low a setting.
This is a very safe device. Keep increasing the
intensity level towards 50 until you feel the
stimulation. You may find that you have to
increase the intensity level as you get used to
the therapy. The aim is not to get to 50 but to
find a setting that is comfortable for you.
Electrode cord is connected to
device.
Disconnect the Electrode cord - the foot-
pads will not work while the Electrode cord is
plugged in.
If, having tried the solutions
above, you still cannot feel the
stimulation:
Test the device by placing one hand across
both footpads at the same time (The heel
end of the footpads is easiest). With your
other hand, and starting from zero, increase
the intensity level until you can feel the
stimulation. If you can feel the stimulation
through your hand then the device is working.
If on 50 you still cannot feel the stimulation
then please contact your authorised dealer.
No power or lights to
Revitive when it is
switched on.
AC Adaptor not switched on at the
electrical socket or AC Adaptor not
plugged into device properly.
Check electrical socket power
is switched on and the AC Adaptor is plugged
into the device correctly. If still not working –
contact your authorised dealer.
Revitive is not vibrating. REVITIVE IS NOT DESIGNED TO VIBRATE.
My legs are aching
aer treatment.
You may have the intensity on
too high a setting and your muscles
are being overworked.
Leave adequate time aer each treatment to
allow the muscles to recover (just like aer
vigorous exercise!). On your next session start
on a lower setting (where you can feel the mild
electrical stimulation and it is comfortable)
and reduce the duration until your muscles
have acclimatized to the stimulation.
Instructions for use
EN
Cleaning and storing Revitive
a Ensure Revitive is switched o.
b Wipe down the foot-pads with a so damp cloth.
Do not clean with chemicals.
Do not immerse Revitive in water.
c Store Revitive in a cool, dry and dust-free location.
Store out of direct sunlight.
d Place Body Pads onto the plastic liner.
e Store the Body Pads in a dry and ventilated location.
Store out of direct sunlight.
You do not have to detach the Body Pads from the Electrode cords to store them.
Replacing the Body Pads
The Body Pads are reusable up to 20-30 applications. If the Body Pads start to lose their adhesiveness, rub a few drops of
water into the surface. Replacement Body Pads can be obtained via www.revitive.com.
Troubleshooting
For more information including Frequently Asked Questions,
please visit www.revitive.com