Below is the recorded cycle PS08_00002.dko opened by Notepad (is installed in every PC).
Using the Notepad program the file can be seen in text format. (
Temp1:= controlling temperature sensor;
Temp2: = verifying temperature sensor; Pres1: = controlling pressure sensor
Cycle report
Owner: Franke
Machine: Deko 190 LC
Serial number: PS08
Cycle Name: NORMAL
Cycle number: 2
Cycle Start Time: 17/11/2010 08:03:39
Time: Temp1: Temp2: Press1:
Cold water flush start
-08:03:54 47 46 0.00
::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::
-08:04:08 47 47 1.07
Cold water flush end
-08:04:15 44 43 0.00
Warm water flush start
-08:04:20 43 43 0.00
::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::
-08:04:44 43 42 0.83
Warm water flush end
-08:04:51 40 39 0.00
Circulation wash start
-08:04:53 40 39 0.00
::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::
-08:05:41 48 48 0.98
Chem set 45ml
T: 48C p:0.99bar
-08:05:45 48 48 0.99
-08:05:48 48 48 0.97
-08:05:50 48 48 0.97
Chem flow OK
-08:05:54 49 48 0.97
-08:05:56 49 48 0.96
::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::
-08:07:32 49 49 0.92
-08:07:36 49 49 0.94
Circulation wash end
T: 49C p:0.99bar
-08:07:39 49 49 0.01
-08:07:42 49 49 0.00
-08:07:44 49 49 0.01
Hot water rinse start
-08:07:50 49 49 0.01
::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::
-08:08:33 56 55 0.93
-08:08:35 56 56 0.93
Hot water rinse end
T: 56C p:0.93bar
-08:08:41 56 56 0.07
Disinfection 90 C start
-08:08:48 56 56 0.01
::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::
-08:12:26 92 93 0.01
Temperature attained:
T: 93C
-08:12:33 94 94 0.01
::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::
-08:13:26 94 94 0.01
Disinfection end
T: 94 °C
-08:13:31 94 94 0.00
Disinfection time 60 s
-08:13:33 93 92 0.00
::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::
-08:13:42 86 85 0.01
Cycle end time: 17/11/2010 08:13:43
Cycle pass: OK