RIGOL Appendix
DG800 Performance Verification Guide
2
DC Offset Accuracy Test
Offset
Setting
Value
Setting
Measurement
Value
Specification Pass/Fail
-2.5V
dc
Frequency:
1kHz
Amplitude:
5Vpp
Impedance:
50Ω
-2.530V
dc
to -2.470V
dc
-1V
dc
-1.045V
dc
to -0.955V
dc
-500mV
dc
-0.550V
dc
to -0.450V
dc
0V
dc
-0.055V
dc
to 0.055V
dc
500mV
dc
0.440V
dc
to 0.560V
dc
1V
dc
0.935V
dc
to 1.065V
dc
2.5V
dc
2.420V
dc
to 2.580V
dc
AC Flatness Test
Frequency
Setting
Value
Setting
Measurement
Value
Calculation
Result
[1]
Specification Pass/Fail
5MHz
Amplitude:
1Vpp
Impedance:
50Ω
±0.1dB
10MHz ±0.2dB
15MHz
[2]
±0.2dB
20MHz
[2]
±0.3dB
25MHz
[2]
±0.3dB
30MHz
[3]
±0.5dB
35MHz
[3]
±0.5dB
Note
[1]
: Calculation result = Measurement value - P
ref
.
Note
[2]
: Only applicable to DG822, DG821, DG832 and DG831.
Note
[3]
: Only applicable to DG832 and DG831.