Chapter 2 Specifications Test RIGOL
DL3000 Performance Verification Manual
Test Procedures
CC programming accuracy refers to the error between the current setting of the
channel and the actual input current value when the electronic load is in CC mode.
CC readback accuracy refers to the error between the readback current value and the
actual input current value when the electronic load is in CC mode.
The following section takes DL3031A as an example to test the CC programming
accuracy and readback accuracy of the DL3000 series.
Note: During the test, please select the corresponding test record form according to
the model of the DL3000 series under test to set the corresponding parameters,
record and calculate the corresponding specifications.
Specification:
Annual Accuracy
[1]
(25 ± 5), ± (% of Output + Offset)
Programming
± (0.05%+0.05%FS
[1]
)
Readback Value
± (0.05%+0.05%FS
[1]
)
Note
[1]
: The accuracy specifications are acquired through calibration at 25 after 1-hour warm-up.
Note
[2]
: FS indicates the full scale.
Procedures:
1. Power off DL3031A. Connect DL3031A, AC power supply, DC power supply, a
current sampling resistor (with a resistance R
M
0.01 Ω), and a digital multimeter
according to Figure 2-1 (Pay attention to the connection method for the current
sampling resistor, the channel input terminals of DL3031A, and the digital
multimeter). Connect the USB DEVICE interface on the rear panel of the
electronic load to the PC via the USB cable.
2. Turn on the AC power supply, and then set its voltage to 230 V, the frequency to
50 Hz. Then, turn on the output.
Note: The voltage setting of the AC power supply should be matched with the
AC voltage selector on the rear panel of DL3031A.
3. Power on DL3031A, and restores all the settings of the electronic load to its
factory default settings.
4. Turn on the DC power supply, and set its output voltage to 8 V and output
current to 62 A.
5. Measure the CC programming accuracy and readback accuracy when the current
range is 60 A.
(1) Measure the CC programming accuracy and readback accuracy when the