EasyManua.ls Logo

Rigol DL3000

Rigol DL3000
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix RIGOL
DL3000 Performance Verification Manual
1
Appendix
Appendix A: Test Record Forms
DL3021 Performance Verification Test Record Form
RIGOL DL3021 Programmable DC Electronic Load
Performance Verification Test Record Form
CC Programming Accuracy and Readback Accuracy
Model
DL3021 (150 V/40 A)
The current range is 40 A.
Specification
Programming Accuracy
± (0.05%+0.02 A)
Readback Accuracy
± (0.05%+0.02 A)
DL3021 Setting
Current
I
L
[1]
Test Result
U
1
I
1
=U
1
/R
M
[2]
I
2
Programming Accuracy: |I
1
- I
L
[1]
|
Readback Accuracy: |I
2
- I
1
|
DL3021 Setting
Current
I
0
[3]
Test Result
U
2
I
3
=U
2
/R
M
[2]
I
4
Programming Accuracy: |I
3
- I
0
[3]
|
Readback Accuracy: |I
4
- I
3
|
The current range is 4 A.
Specification
Programming Accuracy
± (0.05%+0.002 A)
Readback Accuracy
± (0.05%+0.002 A)
DL3021 Setting
Current
I
L
[1]
Test Result
U
1
I
1
=U
1
/R
M
[2]
I
2
Programming Accuracy: |I
1
- I
L
[1]
|
Readback Accuracy: |I
2
- I
1
|
DL3021 Setting
Current
I
0
[3]
Test Result
U
2
I
3
=U
2
/R
M
[2]
I
4

Other manuals for Rigol DL3000