RIGOL Appendix
DL3000 Performance Verification Manual
2
Programming Accuracy: |I
3
- I
0
[3]
|
Readback Accuracy: |I
4
- I
3
|
CV Programming Accuracy and Readback Accuracy
Model
DL3021 (150 V/40 A)
The voltage range is 150 V.
Specification
Programming Accuracy
± (0.05%+0.0375 V)
Readback Accuracy
± (0.05%+0.03 V)
DL3021 Setting
Voltage
U
L
[4]
Test Result
U
1
U
2
Programming Accuracy: |U
1
- U
L
[4]
|
Readback Accuracy: |U
2
- U
1
|
DL3021 Setting
Voltage
U
0
[5]
Test Result
U
3
U
4
Programming Accuracy: |U
3
- U
0
[5]
|
Readback Accuracy: |U
4
- U
3
|
The voltage range is 15 V.
Specification
Programming Accuracy
± (0.05%+0.00375 V)
Readback Accuracy
± (0.05%+0.003 V)
DL3021 Setting
Voltage
U
L
[4]
Test Result
U
1
U
2
Programming Accuracy: |U
1
- U
L
[4]
|
Readback Accuracy: |U
2
- U
1
|
DL3021 Setting
Voltage
U
0
[5]
Test Result
U
3
U
4
Programming Accuracy: |U
3
- U
0
[5]
|
Readback Accuracy: |U
4
- U
3
|

Other manuals for Rigol DL3000