RIGOL Chapter 2 Specifications Test
DL3000 Performance Verification Manual
2-10
Table 2-4 Test Record Form of DL3031A CC Programming Accuracy and Readback
Accuracy
Model
DL3031A (150 V/60 A)
The current range is 60 A.
Specification
Programming Accuracy
± (0.05%+0.03 A)
Readback Accuracy
± (0.05%+0.03 A)
DL3031A Setting
Current
I
L
[1]
Test Result
U
1
I
1
=U
1
/R
M
[2]
I
2
Programming Accuracy: |I
1
- I
L
[1]
|
Readback Accuracy: |I
2
- I
1
|
DL3031A Setting
Current
I
0
[3]
Test Result
U
2
I
3
=U
2
/R
M
[2]
I
4
Programming Accuracy: |I
3
- I
0
[3]
|
Readback Accuracy: |I
4
- I
3
|
The current range is 6 A.
Specification
Programming Accuracy
± (0.05%+0.003 A)
Readback Accuracy
± (0.05%+0.003 A)
DL3031A Setting
Current
I
L
[1]
Test Result
U
1
I
1
=U
1
/R
M
[2]
I
2
Programming Accuracy: |I
1
- I
L
[1]
|
Readback Accuracy: |I
2
- I
1
|
DL3031A Setting
Current
I
0
[3]
Test Result
U
2
I
3
=U
2
/R
M
[2]
I
4
Programming Accuracy: |I
3
- I
0
[3]
|
Readback Accuracy: |I
4
- I
3
|
Note
[1]
: indicates the low current value of the input channel.
Note
[2]
: R
M
= 0.01 Ω.
Note
[3]
: indicates the rated current of the input channel.