Chapter 2 Specifications Test RIGOL
DL3000 Performance Verification Manual
2-15
measurement function and set the range to "Auto". Make sure that the
electronic load works in CV mode. Read the digital multimeter and
record the reading (U
3
) of the digital multimeter.
c) Send the :MEAS:VOLT? command via the remote interface, read and
record the returned voltage value (U
4
) of the query command.
d) Calculate the CV programming accuracy (|U
3
- U
0
|) and the CV
readback accuracy (|U
4
- U
3
|). Compare them with the specified
specifications in Table 2-8. Wherein, U
0
is the rated voltage value of
the input channel.
7. Send the following command via the remote interface to disable the channel
input.
:SOUR:INP:STAT 0
/*Disables the channel input.*/