Chapter 2 Specifications Test RIGOL
DL3000 Performance Verification Manual
2-17
Table 2-6 Test Record Form of DL3021A CV Programming Accuracy and Readback
Accuracy
Model
DL3021A (150 V/40 A)
The voltage range is 150 V.
Specification
Programming Accuracy
± (0.05%+0.0375 V)
Readback Accuracy
± (0.05%+0.03 V)
DL3021A Setting
Voltage
U
L
[1]
Test Result
U
1
U
2
Programming Accuracy: |U
1
- U
L
[1]
|
Readback Accuracy: |U
2
- U
1
|
DL3021A Setting
Voltage
U
0
[2]
Test Result
U
3
U
4
Programming Accuracy: |U
3
- U
0
[2]
|
Readback Accuracy: |U
4
- U
3
|
The voltage range is 15 V.
Specification
Programming Accuracy
± (0.05%+0.00375 V)
Readback Accuracy
± (0.05%+0.003 V)
DL3021A Setting
Voltage
U
L
[1]
Test Result
U
1
U
2
Programming Accuracy: |U
1
- U
L
[1]
|
Readback Accuracy: |U
2
- U
1
|
DL3021A Setting
Voltage
U
0
[2]
Test Result
U
3
U
4
Programming Accuracy: |U
3
- U
0
[2]
|
Readback Accuracy: |U
4
- U
3
|
Note
[1]
: indicates the low voltage value of the input channel.
Note
[2]
: indicates the rated voltage of the input channel.